ES User’s Guide 9-19
Specifications
Table 9-4 System Performance, Corrected, 75Ω, T/R Calibration (continued)
a. These specifications apply for measurements made using “fine” band-
width, no averaging, and at an environmental temperature of 25°±5° C,
with less than 1° C deviation from the calibration temperature.
HP 8712ET/8714ET (75 Ω) without Attenuator
HP 85036B/E (Type-N, 75 Ω) Cal Kit, T/R Calibration
Description
Specification
a
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 48 43
Source Match:
Reflection (One-Port Cal) 41 35
Transmission (Enhanced Response Cal) 41 35
Transmission (Response Cal) 23 19
Load Match 18 15
Reflection Tracking
±0.021 ±0.02
Transmission Tracking:
Enhanced Response Cal
±0.04 ±0.058
Response Cal
±0.11 ±0.2