ES User’s Guide 9-45
Specifications
Features
Measurement
Number of Display Measurements
Two measurement displays are available. Each measurement can have
independent instrument sweep and control parameters (including
frequency settings, IF bandwidth, power level, number of trace points,
averaging, format, and scale). The instrument can display a single
measurement or dual measurements, on a split (two graticules) or
overlaid (one graticule) screen.
Measurement Choices
Narrowband:
ET models: reflection (A/R), transmission (B/R), A, B, R
ES models:
S
11
(A/R), S
22
(B/R), S
21
(B/R), S
12
(A/R), A, B, R
Broadband: X, Y, Y/X, X/Y, Y/R*, power (B*, R*), conversion
loss (B*/R*)
NOTE: X and Y denote external broadband-detector inputs; * denotes
internal broadband detectors.
Formats
rectilinear: log or linear magnitude, SWR
phase, group delay, real and imaginary, Smith chart, and polar
Trace Functions
The analyzer can display current-measurement data, memory data or
current-measurement with memory data simultaneously. Division of
current-measurement data by memory data is also available.