5-6 Chapter5
Language Reference
IEEE Common Commands
Operation Complete
*OPC
*OPC?
Sets bit 0 in the standard event status register to "1" when all pending
operations have finished.
The query stops any new commands from being processed until the
current processing is complete. Then it returns a "1", and the program
continues. This query can be used to synchronize events of other
instruments on the external bus.
Query Instrument Options
*OPT?
This command is not implemented in the HP ESA analyzers as an
IEEE common command. However the function is provided in the ESA
SCPI language reference in the SYSTem subsystem under
:SYSTem:OPTions?.
Recall
*RCL <register>
This command recalls the instrument state from the specified
instrument memory register.
Range: registers are an integer, 2 to 30
Remarks: See also commands :MMEMory:LOAD:STATe and
:MMEMory:STORe:STATe
Front Panel
Access:
File, Recall State
Reset
*RST
This command presets the instrument to a factory defined condition
that is appropriate for remote programming operation. *RST is
equivalent to performing the two commands :SYSTem:PRESet and
*CLS.
NOTE If :SYSTem:PRESet:TYPE is set to USER, then a user preset will be
performed. This presets the analyzer to a state saved by the user. See
:SYSTem:PRESet:TYPE.
Front Panel
Access:
Preset