5-102 Chapter5
Language Reference
SOURce Subsystem
Default Unit: dB
Front Panel
Access: Source, Attenuation Auto Man
Automatic Source Attenuation
:SOURce:POWer:ATTenuation:AUTO OFF|ON|0|1
:SOURce:POWer:ATTenuation:AUTO?
Selects if the source output level attenuator will be set automatically.
Factory Preset
and *RST: On
Front Panel
Access: Source, Attenuation Auto Man
Sets the Output Power
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <ampl>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Specifies the source output power level. Use :SOURce:POWer:SWEep to
set the change in power level across the sweep. See also
:SOURce:POWer:STARt.
Factory Preset
and *RST: −10 dBm
Range: HP ESA E4401B, E4411B: –70 dBm to 3 dBm
HP ESA E4402B, E4403B: –66 dBm to 3 dBm
HP ESA E4404B: –66 dBm to 3 dBm
HP ESA E4405B: –66 dBm to 3 dBm
HP ESA E4407B, E4408B: –66 dBm to 3 dBm
Default Unit: dBm
Front Panel
Access: Source, Amplitude On Off
Sets the Source Output Power Mode
:SOURce:POWer:MODE FIXed|SWEep
:SOURce:POWer:MODE?
Sets the source output to be at a single amplitude (fixed) or to sweep
through a range of power levels.