5-90 Chapter5
Language Reference
SENSe Subsystem
SENSe:HARMonics Subsection
Harmonic Measurement Averages
[:SENSe]:HARMonics:AVERage:COUNt <integer>
[:SENSe]:HARMonics:AVERage:COUNt?
Set the number of averages for the harmonic measurement.
Factory Preset
and *RST: 10
Range: 1 to 1,000
Remarks: This command specifies the number of sweep averages
over which the amplitude of each harmonic and the
total harmonic distortion will be calculated.
Intermediate averaged results will be displayed.
Front Panel
Access:
MEASURE, Harmonic Dist, Meas Setup, Avg Number On
Turn On or Off Harmonic Measurement Averaging
[:SENSe]:HARMonics:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HARMonics:AVERage[:STATe]?
Turn on or off harmonic measurement averaging.
Factory Preset
and *RST: Off
Front Panel
Access:
MEASURE, Harmonic Dist, Meas Setup, Avg Number
Set Harmonic Measurement Sweep Mode
[:SENSe]:HARMonics:SWEeptime[:STATe] AUTO|MAN|1|0
[:SENSe]:HARMonics:SWEeptime[:STATe]?
Set sweep mode per harmonic to auto or manual.
Factory Preset
and *RST: Auto
Remarks: See [:SENSe]:HARMonics:SWEeptime <time>
Front Panel
Access:
MEASURE, Harmonic Dist, Meas Setup, ST/Harmonic
Set Harmonic Measurement Sweep Time
[:SENSe]:HARMonics:SWEeptime <time>
[:SENSe]:HARMonics:SWEeptime?