5-94 Chapter5
Language Reference
SENSe Subsystem
SENSe:POWer Subsection
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator. This value is set at its auto value if input
attenuation is set to auto.
Factory Preset
and *RST: 10 dB
Range: HP ESA E4401B, E4411B: 0 to 60 dB
HP ESA E4402B, E4403B: 0 to 75 dB
HP ESA E4404B: 0 to 75 dB
HP ESA E4405B: 0 to 75 dB
HP ESA E4407B, E4408B: 0 to 65dB
Default Unit: dB
Front Panel
Access:
AMPLITUDE/Y Scale, Attenuation Auto Man
Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the input port attenuator range to be set either automatically or
manually.
On – Input attenuation is automatically set as determined by the
Reference Level Setting.
Off – Input attenuation is manually set
Factory Preset
and *RST: On
Range: HP ESA E4401B, E4411B: 0 to 60 dB
HP ESA E4402B, E4403B: 0 to 75 dB
HP ESA E4404B: 0 to 75 dB
HP ESA E4405B: 0 to 75 dB
HP ESA E4407B, E4408B: 0 to 65dB
Front Panel
Access:
Input, Input Atten