5-84 Chapter5
Language Reference
SENSe Subsystem
SENSe:EBWidth Subsection
Set Emission BW Number of Averages
[:SENSe]:EBWidth:AVERage:COUNt <integer>
Set the emission bandwidth measurement number of averages.
Remarks: This command is used for measurements in the
MEASURE menu.
Front Panel
Access: none
Emission BW Averaging On/Off
[:SENSe]:EBWidth:AVERage[:STATe] OFF|ON|0|1
Turn on or off the channel power measurement averaging.
Factory Preset
and *RST: Off
Remarks: This command is used for measurements in the
MEASURE menu.
Front Panel
Access: Meas Setup, Avg Number On Off
Set Emission BW Span
[:SENSe]:EBWidth:FREQuency:SPAN <freq>
Set the emission bandwidth span. The analyzer span will retain this
value throughout the measurement.
Remarks: This command is used for measurements in the
MEASURE menu.
Front Panel
Access: Meas Setup, EBW Span
Emission BW Trace Max Hold On/Off
[:SENSe]:EBWidth:MAXHold OFF|ON|0|1
Turn on or off the trace max hold trace feature for emission bandwidth
measurements.
Factory Preset
and *RST: Off
Remarks: This command is used for measurements in the
MEASURE menu. Max hold on displays and holds the
maximum responses of a signal; max hold off disables
the feature.