Chapter 5 5-83
Language Reference
SENSe Subsystem
SENSe:DETector Subsection
Type of Detection
[:SENSe]:DETector[:FUNCtion] NEGative|POSitive|SAMPle
[:SENSe]:DETector[:FUNCtion]?
Specifies the detection mode.
Negative peak detection displays the lowest sample taken during the
interval being displayed.
Positive peak detection displays the highest sample taken during the
interval being displayed.
Sample detection displays the first sample taken during the interval
being displayed.
Factory Preset
and *RST: Positive
Front Panel
Access:
Det/Demod, Detector
Det/Demod, Detector, Peak
Det/Demod, Detector, Sample
Det/Demod, Detector, Negative Peak