HT63
EN - 19
6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL CHARACTERISTICS
Accuracy calculated as [%reading + (num. digits*resolution)] at 18°C 28°C <75%HR
DC Voltage
Range Resolution Accuracy Input impedance
Protection against
overcharge
600.0mV 0.1mV
(0.9%reading + 5digits)
>10M
1000VDC/ACrms
6,000V 0,001V
60.00V 0.01V
600.0V 0.1V
1000V 1V
(1.2%reading + 5digits)
AC TRMS Voltage
Range Resolution
Accuracy (*)
Protection against
overcharge
(50Hz1kHz) (1kHz5kHz)
600.0mV 0.1mV
(1.0%reading + 5digits)
(3.0%rdg + 5digits)
1000VDC/ACrms
6,000V 0,001V
60.00V 0.01V
600.0V 0.1V
1000V 1V Not specified
(*) Accuracy specified from 5% to 100% of the measuring range, Input impedance: > 10M
Accuracy PEAK function: 10%rdg, Response time PEAK function: 1ms
Accuracy
function: (1.5%rdg + 5dgt) (f 60Hz), (3.0%rdg + 5dgt) (60Hz f 1kHz), not specified (f: 1 5kHz)
DC Current
Range Resolution Accuracy Protection against overcharge
600.0A 0.1A
(1.5%reading + 5digits)
Quick fuse 800mA/1000V
6000A 1A
60.00mA 0.01mA
600.0mA 0.1mA
10.00A 0.01A Quick fuse 10A/1000V
AC TRMS Current
Range Resolution
Accuracy (*)
Protection against overcharge
(50Hz1kHz) (1kHz5kHz)
600.0A 0.1A
(2.5%rdg + 5digits) (3.5%rdg+5digits)
Quick fuse 800mA/1000V
6000A 1A
60.00mA 0.01mA
600.0mA 0.1mA
10.00A 0.01A Quick fuse 10A/1000V
(*) Accuracy specified from 5% to 100% of the measuring range
Accuracy PEAK function: 10%rdg
4-20mA% reading
Range Resolution Accuracy Correspondence
-25%125%
0.1%
(50digits)
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Diode test
Function Test current Max voltage with open circuit
<1.5mA 2.8VDC