EASYTEST-COMBI519
EN - 12
5.1.2. Country
Move the cursor to Country by means of the arrow keys
(,) and confirm with ENTER. Subsequently, the display
shows the screen which allows selecting the country of
reference, which will influence the LOOP and Ra
measurements.
Select the desired option by means of the arrow keys (,).
To store settings, press the ENTER key, to quit without
confirming the changes made, press the ESC key.
SET
15/10 – 18:04
Europe
Extra Europe
Germany
UK
Norway
USA
Australia/New Zealand
5.1.3. Electrical system
Move the cursor to Electrical system by means of the arrow
keys (,) and confirm with ENTER. Subsequently, the
display shows the screen which allows setting the following
parameters:
Vnom Phase-Neutral or Phase-PE nominal voltage
(110V,115V,120V,127V,133V,220V,230V,240V) to be
used in the calculation of prospective short-circuit current
Frequency system frequency (50Hz, 60Hz)
System type of electric power supply system (TT, TN
or IT)
Contact Volt limit of contact voltage (25V, 50V)
I RCD type of trip-out RCD current visualization (Real,
Nom). With the “Nom” option, the instrument shows the
normalized value of trip-out current (referred to the
nominal current). Example: for RCD type A with
Idn=30mA, the actual value of normalized trip-out current
can be up to 30mA. With the “Real” option, the instrument
shows the actual value of trip-out current by considering
the coefficients indicated by guidelines IEC/EN61008 and
IEC/EN61009 (1.414 for RCD type A, 1 for RCD type AC,
2 for RCD type B). Example: for RCD type A with
Idn=30mA, the actual value of trip-out current can be up to
30mA * 1.414 = 42mA.
RCD/RCCB When selecting the “RCD” option, the
instrument performs the tripping time test with all
multipliers in normal conditions. When selecting the
“RCCB” option, only for 30mA devices, the instrument
performs the tripping time test with x5 multipliers with a
test current of 250mA (type AC) and 350mA (type A).
Isc Factor (only for Norway) possibility to select the
value of the ISC factor (0.01 ÷ 1.00) to be used in the
calculation of prospective short-circuit current.
Select the desired option by means of the arrow keys (,).
To store settings, press the ENTER key, to quit without
confirming the changes made, press the ESC key
SET
15/10 – 18:04
Vnom.
:
230V
Frequency
:
50Hz
System
:
TN
Contact Volt
:
5
V
I RCD
:
Nom.
RCD/RCCB
:
RCD
Isc Factor
:
0.75