6.4.1. Anomalous situations
1. If the instrument detects a voltage higher than 3V at its
terminals, it does not perform the test, it emits a prolonged
acoustic signal and displays a screen like the one to the
side.
Lo 15/10 – 18:04
R
Itest
=
=
- - -
- - - A
Vin > 3V
0.500
Conn.
Lim. INFO
2.
If the instrument does not detect the Lo 10A accessory, it
displays a screen like the one to the side. Check the
connections with the accessory.
Lo 15/10 – 18:04
R
Itest
=
=
- - -
- - - A
Accessory not detected
0.500
Conn.
Lim. INFO
3.
The instrument shows on the display the message "NOT
OK" in case of a positive result (value lower than the set
limit threshold) but with test current lower than 10A, as
indicated in the screen like the one to the side.
Lo 15/10 – 18:04
R
Itest
=
=
0.119
8.05 A
NOT OK
0.500
Conn.
Lim. INFO