6.4. LO: CONTINUITY OF PROTECTIVE CONDUCTORS WITH 10A
This function allows measuring the resistance of protective and equipotential conductors
with a test current >10A by using the Lo 10A optional accessory connected to the
instrument through the C2050 cable. The optional accessory Lo 10A must be directly
powered by the mains on which measurements are performed. For detailed information,
please refer to the user manual of the relative accessory
CAUTION
The instrument can be used for measurements on installations with
overvoltage category CAT IV 300V to earth and max 415V between
inputs.
We recommend holding the alligator clip respecting the safety area
created by the hand protection (see § 4.2).
Check that no voltage is present at the ends of the item to be tested
before carrying out a continuity test.
The results may be influenced by the presence of auxiliary circuits
connected in parallel with the item to be tested or by transient currents.
Continuity test is carried out by supplying a current higher than 10A in
case the resistance does not exceed ca. 0.7 (including resistance of
the test cables). The 4-wire method allows extending the test leads
without any preliminary calibration.
1.
Press the MENU key, move the cursor to Lo in the main
menu by means of the arrow keys (,) and confirm
with ENTER. Subsequently, the instrument displays a
screen similar to the one reported here to the side.
Lo 15/10 – 18:04
R
Itest
=
=
- - -
- - - A
0.500
Lim. INFO
2. Use the , keys to modify the parameter value.
Lim this virtual key allows the selection of the maximum limit to consider the
measured value correct. It is possible to set a limit included in the range: 0.003 ÷
0.500 in steps of 0.001
3.
Connect the Lo 10A accessory to the mains
(230/240V – 50/60Hz) and check that the green LED
lights up. Connect the accessory to the instrument
through the C2050 cable. Subsequently, the “Conn.”
message is shown indicating the correct detection by the
instrument.
Lo 15/10 – 18:04
R
Itest
=
=
- - -
- - - A
0.500
Conn.
Lim. INFO

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HT EASYTEST COMBI519 and is the answer not in the manual?

HT EASYTEST COMBI519 Specifications

General IconGeneral
BrandHT
ModelEASYTEST COMBI519
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals