9. Press the GO/STOP key to start the test. In case no error
conditions occur, the instrument starts a series of
continuous measurements for the entire duration of the
set Timer, emitting a short sound every 3s, and
showing alternatively the messages “Measuring…” and
Please wait…” as shown in the screen to the side. In this
way, the operator can move from one point to another of
the place in which measurement is being carried out.
RPE 15/10 – 18:04
R 0.23
Itest 209 mA
T 11 S
Please wait…
STD 2.00 12s 0.06
MODE Lim. Time
><
10. At the end of measurement, the instrument provides the
maximum value among all those of the partial
measurements carried out. If the result is lower than the
set maximum limit value, the instrument shows the
message “OK” (value lower or equal to the set limit
threshold), otherwise it displays the message “NO OK
(value higher than the set limit threshold) as shown in the
screen to the side.
11. Press the SAVE key to store the test result in the
instrument’s memory (see § 7.1) or the ESC/MENU key to
exit the screen without saving and go back to the main
measuring screen.
RPE 15/10 – 18:04
R 0.54
Itest 209 mA
T 0 S
OK
STD 2.00 12s 0.06
MODE Lim. Time
><