Accuracy is indicated as ±[%reading + (number of digits*resolution)] at 23°C±5°C, <80%RH
DC Voltage @ OPC
Range [V] Resolution [V] Accuracy
3.0 1500.0
0.1
(1.0rdg + 2dgt)
Minimum VPN voltage to start the test: 15V
DC Current @ OPC
Range [A] Resolution [A] Accuracy
0.10 40.00
0.01
(1.0rdg + 2dgt)
DC Voltage @ STC
Range [V] Resolution [V] Accuracy
3.0 1500.0
0.1
(4.0rdg + 2dgt)
DC Current @ STC
Range [A] Resolution [A] Accuracy
0.10 40.00
0.01
(4.0rdg + 2dgt)
10.2. GENERAL CHARACTERISTICS
Reference guidelines
Instrument safety: IEC/EN61010-1, IEC/EN61557-1
EMC: IEC/EN61326-1
Safety of measuring accessories: IEC/EN61010-031
Measurements: IEC/EN62446 (IVCK)
IEC/EN61557-2 (M)
IEC/EN61557-4 (RPE)
Insulation: double insulation
Pollution level: 2
Measurement category: CAT III 1000VAC, CAT III 1500VDC to earth,
Max 1000VAC, 1500VDC between inputs
Display, memory and PC interface
Type of display: LCD custom, 240x240pxl, with backlight
Saved data: max 999
Internal database: max 64 modules
PC interface: optical/USB and WiFi
Interface with SOLAR03: Bluetooth connection (max distance 100m)
Power supply
Battery type: 6x1.5V alkaline type AA LR06 or
6x1.2V rechargeable batteries NiMH type AA LR06
Low battery indication: symbol “ ” on the display
Battery duration (@Temp = 20°C): RPE: >500 Tests (RPE ≥ 0.1)
GFL, M: >500 tests (Riso≥1kxVTest)
IVCK: >500 tests (no SOLAR03)
Auto power off: after 5 minutes' idling
Mechanical characteristics
Size (L x W x H) 235 x 165 x 75mm
Weight (batteries included): 1.2kg
Mechanical protection: IP40