EasyManua.ls Logo

Humeau LIGHTNING MVP ICON - ATP Swabs to Target Widget; ATP Re-Tests Widget

Humeau LIGHTNING MVP ICON
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
From “Device” screen select “Calibration”.
Press the “Positive Control” button under the Verificationmenu in ATP Calibration
mode.
Open the ATP Positive Control vial.
Insert a LIGHTNING MVP ICON Surface Sampling Device and completely dissolve the
positive control pellet onto the tip of the swab.
Note: Take care to ensure the swab tip does not touch any surface other than the inside
of the vial.
Activate the swab, insert into the chamber with swab chamber facing left to right, and
close the chamber door.
Press “Verify” to run the verification.
The result for a positive control should read between 3.1 4.3 Zones.
Results greater than 4.3 indicate a potential contamination of either the swab or positive
control. Re-test in duplicate and average the two results.
Between 2.6 3.0 Zones, more thorough swabbing of the vial may be required.
Below 2.6 Zones, ensure the device was properly activated. If device was activated
properly, it may indicate a problem with the instrument, device, or positive controls.
Repeat the test with a new vial of positive control and new sampling device.
If verification fails again, calibrate the instrument and repeat verification with a new vial
of positive control and sampling device.
If still out of range, contact BioControl Technical Support at 800.245.0113.
Negative Control Procedure
A negative control is an unused but activated swab.
Press the “Negative Control” button under the Verification menu in ATP Calibration.
Activate the swab, insert into the chamber with the swab chamber facing left to right,
and close the chamber door.
Press “Verify” to run the verification.
The result for a negative swab should read between 1.0 2.5 Zones.
LIGHTNING MVP ICON Instrument Verification
To verify if the instrument is reading correctly, the luminescent and non-luminescent calibrators
are used.
The luminescent calibrator consists of a
14
C radioactive source which emits a very low level (5x10
5
d.p.m. or 0.225 microcuries) of low-energy β radiation in a plastic scintillating matrix. This matrix
is configured to provide a constant light output at a sufficiently high level in order to perform
consistent verification and calibration over the shelf-life of the calibrator.
The non-luminescent calibrator provides a reliable low-end or “dark” source reading.
Luminescent Calibrator Procedure
Under the “Verification” menu in ATP Calibration, select “Luminescent Calibrator”.
Insert the luminescent calibrator and close the chamber door.
Press “Verify”.
In a properly calibrated MVP ICON, the calibrator should read within 0.1 Zones of the zone value
displayed on the calibrator’s label.
21

Table of Contents