EasyManuals Logo

IKONIX Associated Research Hypot III User Manual

Default Icon
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
35
At the Cal-Alert Warning screen “Show screen again?” will be displayed along with the
options Yes or “No.” The question must be answered by toggling between the words
“Yes” and “No” using the corresponding soft key and then pressing the OK soft key.
Selecting “No” will disable the Cal-Alert function. Selecting “Yes” will exit the Cal-Alert
Warning screen and go to the standard introduction screen without disabling the Cal-
Alert function.
If security is enabled this screen will not appear.
It is possible to quickly bypass the security of the Cal-Alert by powering up the
instrument while pressing the soft key that is second from the top of the instrument.
This will temporarily start the instrument in an unsecured state. The Cal-Alert Warning
screen will be displayed and Cal-Alert may be disabled by selecting “No” and then
pressing the OK soft key. This type of security override only affects the Cal-Alert
screen. It is not necessary to restart the instrument to activate the security setting.
4.2.3. Default System Parameters
Setup Sys.
PLC Remote
OFF
Single Step
OFF
Alarm
5
Contrast
5
Results
Last
Lock
OFF
Mem Lock
ON
Smart GFI
ON
4.2.4. Memory, Step, and Default System Parameter Restore
Restoring the instrument’s System parameters will overwrite
all memories and steps with ACW default parameters!
Press the two bottom soft keys and power the instrument at the same time, then press
the soft key next to the word “Yes”. All memories and steps will be loaded with the
ACW default parameters and the system parameters will be set to the factory defaults.
The following System parameters will not be affected by the system initialization:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IKONIX Associated Research Hypot III and is the answer not in the manual?

IKONIX Associated Research Hypot III Specifications

General IconGeneral
BrandIKONIX
ModelAssociated Research Hypot III
CategoryMeasuring Instruments
LanguageEnglish