15
3. SPECIFICATIONS AND CONTROLS
3.1. 3705, 3765, 3770 Functional Specifications
INPUT
Voltage
115/230 VAC 10, user selectable
Frequency
50/60 Hz 5%
Fuse
3.15 Amp, fast acting 250VAC
DIELECTRIC WITHSTAND TEST MODE
Output Rating
5000V @ 20 mA AC
6000V @ 7.5 mA DC
Voltage Setting
Range:
0 5.00KV AC
0 6.00KV DC
Resolution:
0.01KV
Accuracy:
(2 of setting 5V)
(Adjustable during operation.
Disabled when key lockout is active.)
Maximum Limit AC
Range:
0.00 20.00 mA
Resolution:
0.01 mA
DC
Range:
0 7500 A
Resolution:
1A
Accuracy:
AC and DC (2 of setting 2 counts)
Minimum Limit AC
Range:
0.000 9.999 mA
Resolution:
0.001 mA
DC
Range:
0.0 999.9 A
Resolution:
0.1 A
Accuracy:
AC and DC (2 of setting 2 counts)
Arc Detection
Range:
0 9, 0 disabled
Failure Detector
Audible and Visual
Ground Fault Interrupt
GFI Trip Current: 450A max (AC or DC)
HV Shut Down Speed: <1ms
Voltage Display
Range:
0.00 5.00KV AC
0.00 6.00KV DC
Resolution:
0.01KV
Accuracy:
(2 of reading 0.01KV)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IKONIX Associated Research Hypot III and is the answer not in the manual?

IKONIX Associated Research Hypot III Specifications

General IconGeneral
BrandIKONIX
ModelAssociated Research Hypot III
CategoryMeasuring Instruments
LanguageEnglish