54
7.2.4. System Parameter Editing Commands and Companion Queries
These commands are used to modify the system parameters for the instrument.
These
commands require a parameter value to be included with the command.
7.2.5. Query Commands
These query commands will retrieve data from the instrument. These commands
include functions for retrieving test data, test results and remote hardware.
Associated Research Inc., Model
Number, Serial Number, Firmware
Revision
Data from test in Process
Response:
SALL
memory+step,type,p1,p2,p3…
*IDN?
Read the instrument identification string. Company = Associated Research Inc.
TD?
Read the active data being displayed on the LCD display while the test is in process. It
will also read the last data taken when the test sequence has completed. Each
parameter is separated by commas and includes step number, test type, test status,
and metering. The syntax for this command response is {memory - step, test type,
status, meter 1, meter 2, meter 3}. ACW test displays 4 meters. Each meter will
contain both the value and the units. In the case of DCW current where both uA and