62
9.2.2. AC Hipot Verification
From the Verification screen, press the ” and “soft keys to select “AC Hipot.
Then press the Select soft key. An instructional prompt will now be displayed. The
prompt will appear as follows:
Follow the instructions given in the prompt and then press TEST to begin the
verification process.
At the end of the verification process a message will appear indicating the outcome of
the process. If the instrument passes the verification (test failure, indicating the fail
detectors are working properly) the RESET button will illuminate, the alarm will sound
and the following message will appear:
If the instrument fails the verification (test pass, indicating the fail detectors are not
working properly) the following message will appear: