44
NOTE: For multi-step tests, use the “” and “” soft keys to scroll through the
results.
5.5. Error Messages
While performing tests a number of messages will be displayed to indicate the test
state or test results. These messages are displayed in the status area of the screen
located on the first line at the top of the LCD. These messages are also used on other
screens where test results are displayed.
Abort: This message appears on the display if the test in process is aborted with the
RESET button or remote RESET control.
Max-Fail: This message appears on the display if the measured reading exceeds the
Max Lmt current setting of the AC/DC Withstand test or Max Lmt resistance setting of
the Insulation Resistance test.
Min-Fail: This message appears on the display if the measured reading drops below
the Min Lmt current setting of the AC/DC Withstand test or Min Lmt resistance setting
of the Insulation Resistance test.
Cont. Max: This message appears on the display during the AC/DC Withstand test if
the measured reading exceeds the Max Lmt continuity resistance setting (Continuity
function must be turned ON).
Cont. Min: This message appears on the display during the AC/DC Withstand test if
the measured reading drops below the Min Lmt continuity resistance setting
(Continuity function must be turned ON).