49
If the Remote Interlock contacts are opened during a test, the pop-up message
will be displayed and the test will abort. The hardware and has been configured
to provide the interlock connections on pins 4 and 5 of the Remote Interface,
Signal Input port. The instrument can still be used without the external interlock
device as long as the Interlock Disable Key (38075 provided with unit) is plugged
into the Remote Interface, Signal Input port. If there is nothing connected to the
Remote Interface, Signal Input port to provide a connection to the Remote
Interlock, the instrument will not perform tests.