52
Select
step number 1-3
ACW
Step Add ACW
test
DCW
Step Add DCW
test
IR
Step Add IR test
SALL<memory+step,type,p1,p2,p3…>
Add all
Parameters of a
Test
MSS<memory number step number>
Load a file by memory and step number. The first digit is the memory number and the
second digit is the step number. Example: MSS <11> = memory 1 step 1
ACW, DCW, IR
These commands add the appropriate test type within the memory at the step location
that has been selected. The parameters of the previous test type will be deleted and
the default values for the new test type will be recalled. If the same test type is
selected that already exists, the default values will replace the previous parameters.
SALL <memory+step, type, p1, p2, p3… >
This command edits all parameters in a memory and step. Parameters will be edited
at the step location that has been selected. See the command summary tables to see
the specific test type for each of these commands. Example:
SALL<01,ACW,1.24,10.00…>
The parameter <memory+step> indicated the memory and step number. The
parameter <type> indicated the test type. The values ACW, DCW, or IR must be used.
The parameters <p1,p2> etc. indicate the individual settings for each test. All
parameters must be included with the command and should appear in the same order
that is shown in the table below. Also, like the individual parameter editing commands,
the unit should not be included with the value. Only the numeric value should be
included in the command string. The list of parameters can also be found in the
default parameters section of the manual, or refer to the Test Parameter Editing
commands section for the proper values. The parameter values should use complete
text and not use the coded values that are associated with the individual parameter
setting commands such as "ON" and "OFF" for toggle field.
ACW
DCW
IR
1
Voltage
Voltage
Voltage
2
Max Limit
Max Limit
Max Limit
3
Min Limit
Min Limit
Min Limit
4
Ramp Up
Ramp Up
Ramp Up
5
Dwell
Dwell
Delay
6
Ramp Down
Ramp Down
Ramp Down
7
Arc Sense
Arc Sense
Connect
(ON/OFF)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IKONIX Associated Research Hypot III and is the answer not in the manual?

IKONIX Associated Research Hypot III Specifications

General IconGeneral
BrandIKONIX
ModelAssociated Research Hypot III
CategoryMeasuring Instruments
LanguageEnglish