OtoRead™ - Instructions for Use - US Page 23
Example of a pass result.
If an error message appears during testing or if one or more of the OAE bands or points has a checkmark
(meaning detected), the probe integrity test has failed. This could indicate that:
• There is wax or debris in the probe tube and replacing it is required.
• The probe was not placed in the infant ear simulator correctly, or,
• The probe’s calibration needs to be checked.
Check and replace the probe tube if necessary and retry the test. If the probe test fails a second time, the
probe should not be used to test on patients. Contact your local service technician for assistance.
NOTE: It is possible for system distortion to appear at levels below -10dB SPL. Ensure that the MIN OAE
LEVEL is set to -10 dBSPL on the device for the protocols in use before conducting the probe integrity test.