OtoRead - Instructions for Use - US Page 23
Example of a pass result.
If an error message appears during testing or if one or more of the OAE bands or points has a checkmark
(meaning detected), the probe integrity test has failed. This could indicate that:
There is wax or debris in the probe tube and replacing it is required.
The probe was not placed in the infant ear simulator correctly, or,
The probe’s calibration needs to be checked.
Check and replace the probe tube if necessary and retry the test. If the probe test fails a second time, the
probe should not be used to test on patients. Contact your local service technician for assistance.
NOTE: It is possible for system distortion to appear at levels below -10dB SPL. Ensure that the MIN OAE
LEVEL is set to -10 dBSPL on the device for the protocols in use before conducting the probe integrity test.