D-0116684-B – 2020/03
OtoRead™ - Instructions for Use - US Page 56
Too High
For a DP test, the level of the calibration tone is too high.
User should refit the probe in the patient’s ear and retry the test
Replace the probe tube
Too Low
For a DP test, the level of the calibration tone is too low.
User should refit the probe in the patient’s ear and retry the test
Replace the probe tube
Error
Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test
Cycle instrument power
Saved tests are within 5 tests of the maximum limit.
Print or transfer test results to the PC to avoid interruption in testing
The maximum saved test limit is reached.
The user will need to clear the memory before any additional tests can be
The battery charge level is too low for operation.
The user must charge the battery before additional tests can be performed
Indicates a problem with the printer.
Check the printer status
Reset the printer or cycle the printer power
Indicates the paper has run out.
Replace the paper roll
The clock is checked during power on to ensure it has not lost time and been
reset. In the case of clock reset, this message is shown.
The user should set the correct date/time
Indicates that calibration of instrument is due. The message will appear upon
the calibration due date set in the device. A daily reminder will appear during
device startup.
Arrange calibration of the instrument
Contact Interacoustics or your local distributor for service if any of the above problems persist.