EasyManua.ls Logo

Interacoustics OtoRead - Page 77

Interacoustics OtoRead
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0116684-B 2020/03
OtoRead - Instructions for Use - US Page 71
7.3 Appendix C: Configurations and test protocols
DPOAE protocols
Protocolname
# of
Freq.
F2 Freq. [kHz]
L1/L2
Averaging
Time [s]
Pass SNR
[dB]
# Passing
Freq. for
Test Pass
Screening
DP 2s
4
2, 3, 4, 5
65/55
2
6
3
DP 4s
4
2, 3, 4, 5
65/55
4
6
3
Clinical
DP 2.0-5.0
4
2, 3, 4, 5
65/55
4
6
3
DP 1.5-6.0
6
1.5, 2, 3, 4, 5, 6
65/55
4
6
0
DP 1.6-8.0
12
1.6, 2, 2.5, 3.2,
3.6, 4, 4.5, 5,
5.6, 6.3, 7.1, 8
65/55
4
6
0
DP 1.5-12
12
1.5, 2, 3, 4, 5, 6,
7, 8, 9, 10, 11,
12
65/55
4
6
0
(Diagnostic version also includes DP 4s screening protocol)
Grey fields are customizable fields:
L1/L2
: 40 to 70 dBSPL
Average time
: 0.5, 1, 2 or 4 sec.
Pass SNR
: 3 to 10 dB
Passing Freq. for Test Pass
: 1 to 12
TEOAE protocols
Protocolname
# of
Freq.
Freq. [kHz]
Averaging
Time [s]
Pass SNR
[dB]
# Passing
Freq. for
Test Pass
Screening
TE 32s
6
1.5, 2, 2.5, 3, 3.5, 4
32
4
3
TE 64s
6
1.5, 2, 2.5, 3, 3.5, 4
64
4
3
Clinical
TE 1.5 4.0
6
1.5, 2, 2.5, 3, 3.5, 4
64
4
3
TE 0.7 4.0
6
0.7, 1, 1.4, 2, 2.8, 4
64
4
0
(Diagnostic version also includes TE 64s screening protocol)
Grey fields are customizable fields:
Average time
: 4, 16, 32 or 64 sec.
Pass SNR
: 3 to 10 dB
Passing Freq. for Test Pass
: 1 to 6

Table of Contents