The system will exercise the cell based on the number of cycles indicated by the ExerciseCycle parameter
set in the AUTO.CFG parameter secon of the WgoldCfg program. The system will rst apply a load
approximang the percent load indicated by the InialLoad parameter. The output of the standard is
compared against a nominal output based on the set point le sengs. The output of the DUT is compared
against a nominal output based on the values found in the GOLDINT.LMT or GOLDUSR.LMT les. If either
reading is found to be out of tolerance, a FAIL SAFE prompt will be displayed and the calibraon cycle
aborted. Refer to the AUTO.CFG le parameters in the WGoldCfg program.
Calibrate DUT.
The set point box located in the lower le corner should now read 0. When ready for the inial zero to be
taken, select the START buon. The set point number will change to a highlighted color and aer a few
seconds a beep will be heard indicang the reading has been taken. The next desired set point will then be
displayed in the set point box.
Increase the load unl the STANDARD meter is within the limits dened in the ADC.CFG le. (Automated
systems will automacally increase the load as needed) The set point number will change to a highlighted
color. Connue holding steady unl a beep is heard and the next set point is displayed. The nal set point
number is for zero load will equal inial zero reading taken on the standard. All load should be removed
and the system will record the return zero.
The gauge display can be used as a reference. When the set point meter display matches the reference
standard meter display, the gauge indicator will be centered.
Pressing the ESC key at any me during the calibraon will abort the calibraon and return to step to the
beginning of step 10.
Performance Analysis.
The system will now display the performance results. Output, Nonlinearity, Hysteresis and SEB parameters
are compared against limits found in the GOLDINT.LMT or GOLDUSR.LMT les. A High Or Low displayed to
the right of these parameters indicates this parameter exceeds the limits found.
The following buons are displayed in the lower right corner.
Displays results for a mulple bridge transducer.
SHUNT CAL
Select if shunt cal reading(s) are desired.
The program defaults to a value to provide Auto shuntcal output of about 70% of calibrated full scale on
a 350 ohm bridge if possible. If an external resistor is to be used, select Manual and enter its value in the
edit eld.
PRINTER
Generates a hard copy of the performance results.
A menu will be displayed to select the desired report to print. These reports are created using the Report
Designer found on the Tools menu from the main screen.
This will produce a load vs. error plot. Before prinng, it is recommended that the color be changed to
black and white.