The system will exercise the cell based on the number of cycles indicated by the ExerciseCycle parameter
set in the AUTO.CFG parameter secon of the WgoldCfg program. The system will rst apply a load
approximang the percent load indicated by the InialLoad parameter. The output of the standard is
compared against a nominal output based on the set point le sengs. The output of the DUT is compared
against a nominal output based on the values found in the GOLDINT.LMT or GOLDUSR.LMT les. If either
reading is found to be out of tolerance, a FAIL SAFE prompt will be displayed and the calibraon cycle
aborted. Refer to the AUTO.CFG le parameters in the WGoldCfg program.

Calibrate DUT.
The set point box located in the lower le corner should now read 0. When ready for the inial zero to be
taken, select the START buon. The set point number will change to a highlighted color and aer a few
seconds a beep will be heard indicang the reading has been taken. The next desired set point will then be
displayed in the set point box.
Increase the load unl the STANDARD meter is within the limits dened in the ADC.CFG le. (Automated
systems will automacally increase the load as needed) The set point number will change to a highlighted
color. Connue holding steady unl a beep is heard and the next set point is displayed. The nal set point
number is for zero load will equal inial zero reading taken on the standard. All load should be removed
and the system will record the return zero.
The gauge display can be used as a reference. When the set point meter display matches the reference
standard meter display, the gauge indicator will be centered.
Pressing the ESC key at any me during the calibraon will abort the calibraon and return to step to the
beginning of step 10.

Performance Analysis.
The system will now display the performance results. Output, Nonlinearity, Hysteresis and SEB parameters
are compared against limits found in the GOLDINT.LMT or GOLDUSR.LMT les. A High Or Low displayed to
the right of these parameters indicates this parameter exceeds the limits found.
The following buons are displayed in the lower right corner.

Displays results for a mulple bridge transducer.
SHUNT CAL
Select if shunt cal reading(s) are desired.
The program defaults to a value to provide Auto shuntcal output of about 70% of calibrated full scale on
a 350 ohm bridge if possible. If an external resistor is to be used, select Manual and enter its value in the
edit eld.
PRINTER
Generates a hard copy of the performance results.
A menu will be displayed to select the desired report to print. These reports are created using the Report
Designer found on the Tools menu from the main screen.

This will produce a load vs. error plot. Before prinng, it is recommended that the color be changed to
black and white.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface GOLD STANDARD and is the answer not in the manual?

Interface GOLD STANDARD Specifications

General IconGeneral
BrandInterface
ModelGOLD STANDARD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals