EasyManuals Logo

Interface GOLD STANDARD User Manual

Interface GOLD STANDARD
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Page 26 of 45

CALIBRATION SET POINT FILE CREATION
When a new calibraon sequence is desired or a calibraon standard is recalibrated, a new calibraon set
point le is required. Following is the instrucon for creang a le. When all of the required data has been
entered in the Device Under Test informaon screen, select either TENSION or COMPRESSION for the mode of
calibraon. The set point selecon screen will now be displayed (See gure 9). There are 3 dierent methods
of le creaon. Each is explained.
Using mV/V values from the standard calibraon cercate.
Follow these steps. (A typical 5 point, 5Klbf calibraon le is illustrated as an example).
1. Enter into the load column each of the ve (5) ascending and one (1) descending load points as they are
taken from the calibraon sheet of the standard. Make certain to use the same units as the standard
calibraon sheet (Klbf in the example).
2. Press the Enter key to get to the top of the next column. Enter the mV/V value for each set point from
the standard calibraon sheet. Make certain the entered values are for the desired mode, tension or
compression, and that the polarity of each value as listed is preserved.
3. When all of the values have been entered, click the Curve Fit Plot buon. The graph should show a
smooth curve with all points falling on the curve. If not, a keying error has been made and must be
corrected. If the graph is OK, click the Exit buon and then the Connue buon. The le can then be
saved under the desired le name.

 Using coecients from the standard calibraon cercate.
1. This method is especially useful for creang set points that are not listed on the standard calibraon
cercate. Enter the desired loads in the load column. Make certain they are in the same units as the
standard calibraon cercate.
2. Select the polynomial degree of t.
3. Enter the coecient values from the calibraon cercate for the appropriate mode.
4. Click the “Compute Set Points” buon. The mV/V column will be completed automacally.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface GOLD STANDARD and is the answer not in the manual?

Interface GOLD STANDARD Specifications

General IconGeneral
BrandInterface
ModelGOLD STANDARD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals