7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Page 21 of 45

This will display the ed curve output and error as well as the curve coecients and standard deviaon. A
Printer buon allows a hard copy report to be printed. These reports are created using the Report Designer
found on the Tools menu from the main screen.

This will produce a load vs. ed curve error plot. Before prinng, it is recommended that the color be
changed to black and white.
REPEAT
Repeats the calibraon cycle beginning with step 10.
INFO SCREEN
Returns to step 2.
PERFORM CREEP

Select Perform Creep buon on toolbar or from pull down Main Menu.

Enter all DUT informaon.

Select DUT mode (Tension or Compression)

Enter Set Point Informaon. Creep test only uses the full scale value entered.

Select Standard Mode (Tension or Compression). Normally this is the same as the DUT Mode.

Select the Connue Buon. If the set points were newly entered, or an exisng set point le has been
edited, a screen will be displayed asking to Save? Selecng Yes will then display a Select File window where
an exisng le can be selected to overwrite or a new le name entered.

Enter desired Creep Delays in seconds. Selecng the Default buon will use default creep sengs. The
Select buon allows loading of previously entered and saved Creep delay sengs.

Select the Connue Buon.

The set point box located in the lower le corner should now read 0. When ready for the inial zero to
be taken, select the START buon. The set point number will change to a highlighted color and aer a
few seconds a beep will be heard indicang the zero reading has been taken. Increase the load unl the
STANDARD meter is within the limits dened in the ADC.CFG le.
(Automated systems will automacally increase the load as needed) All load should be removed and the
system will record the return zero. A mer will begin counng down from the maximum delay entered in
Step 8. When the mer reaches zero the system will prompt to remove load and click Connue Buon.
The mer will begin counng down as readings are recorded per the delays entered in Step 8. Once all
readings are recorded the Creep results are displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface GOLD STANDARD and is the answer not in the manual?

Interface GOLD STANDARD Specifications

General IconGeneral
BrandInterface
ModelGOLD STANDARD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals