Page #11 background image
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Page 11 of 45
GS-SYS




4 column loading frame
150mm (6 inch) stroke
Alignment swivel coupler/slack adapter
Static calibration in tension or compression
Top swivel coupler thread – 2”-12 male
Bottom actuator thread – 2”-12 female
21MPa (3000 psi) hydraulic power unit
25,000 lbf. load capacity/24.4 to 30.4 inches working length
55,000 lbf. load capacity/24.4 to 30.4 inches working length
55,000 lbf. load capacity/38.4 to 44.4 inches working length
100,000 lbf. load capacity/38.4 to 44.4 inches working length
208/230 VAC, single phase, 50/60 hertz
Power – 3.5 kVA
Amps (continuous) – 16
Minimum protection circuit current – 30 Amps

Normal operaon 2600 psi
Maximum pressure 3000 psi
10 gal ISO 32 hydraulic oil.

Moog Direct Drive Valve D633-303B-R02K01M0NSM2
Flow Rang: 6 lpm ( 1.6 gpm)
Seal Material: NBR
Minimum Supply Pressure: 15 bar ( 220 psi)
Maximum Supply Pressure: 350 bar (5000 psi)
Filtraon for Normal Operaon: 10 micron absolute
Fluid Cleanliness Level: ISO 4406 15/13/10 or NAS 1638 Class 4
Coil Resistance: 25 Ω
Command Signal: ± 600mA
Power Consumpon: 9 W (at I = 600 mA and R = 25 Ω)

Delta Computer: RMC75E-AA2 with AP2 and D8 expansion modules
Primary Communicaon Type: Ethernet
Monitor Port: USB
Control Loop Time: User selectable from 0.5 to 4 ms
Voltage: +24 VDC ± 20%
Current: Typical for base model 200 mA at 24 VDC
DC-DC Converter Isolaon: 500 VAC input to controller
Operang Temperature: +32 to +140 °F
(L6-30R Connector on system)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface GOLD STANDARD and is the answer not in the manual?

Interface GOLD STANDARD Specifications

General IconGeneral
BrandInterface
ModelGOLD STANDARD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals