I-Lift™
SYMPTOMSFAULTSSOLUTION
Hand-controloractuatorconnectorloose.
Connecthandcontroloractuatorconnector.
Ensureconnectorsareseatedproperlyand
fullyconnected.
Batterylow.
Chargebatteries.Referto5.7Chargingthe
Battery,page24.
REDemergencystopbuttonpressedIN.RotateREDemergencystopbutton
CLOCKWISEuntilitpopsout.
Batterynotconnectedproperlytocontrol
box.
Reconnectthebatterytothecontrolbox.
Referto5.7ChargingtheBattery,page24.
Theconnectingterminalsaredamaged.
Replacethebatterypack.Referto5.7
ChargingtheBattery,page24.
Electricactuatorfailstoliftorlegsfailtoopen
whenbuttonispressed.
Boomorlegactuatorinneedofserviceor
loadistoohigh.
Referto8.9ReplacingtheElectricActuator,
page42.or8.10ReplacingtheLegActuator,
page43.ContactyourDealer.
Unusualnoisefromactuator.Actuatoriswornordamagedorspindleis
bent.
Referto8.7ReplacingaManual/Hydraulic
Pump,page40,8.9ReplacingtheElectric
Actuator,page42or8.10ReplacingtheLeg
Actuator,page43.ContactyourDealer.
Boomwillnotlowerinuppermostposition.Boomrequiresaminimumweightloadto
lowerfromtheuppermostposition.
Pulldownslightlyontheboom.
Boomwillnotlowerduringapower
retraction.
Shoulderboltatthejunctionoftheboomand
mastmaynotbeproperlyinstalled.
Referto8.11CheckingandTighteningMast
PivotBolt,page44.
Hydraulicpumpdriftswhenunderload.Controlvalvemaybeoutofadjustment.Referto8.8AdjustingthePumpControl
Valve,page41.
Ifproblemsarenotremediedbythesuggestedmeans,pleasecontactyourdealerorInvacare.