RS-485 Digital Communications
116
Rev. L6 Feb 2017 Modline 5
Returns the temperature the unit is reading, followed by a comma, then the current status (see the
ST command for details), another comma, and then the current attenuation percent (expressed as a
number from 0 to 100)
Range: N/A
TS TS Temperature and Status (Read Only)
Returns the temperature the unit is reading followed by a comma, and then either a 0 (no alarm
conditions), or a number with the same format as the “ST” Command.
Range: measured temperature in degrees C or F and 32768 to 32767.
ST - System Alarm and Measurement Condition Status (Read Only)
The Sensor returns a message indicating the current measurement condition status and the status of
alarms.
Range: Not Applicable
Values: Bit mapped values for alarm and measurement status.
Value
Condition
Equivalent Display or Error Code
0
None to Report
None
1
Out of calibration
Same as Error Code X102
2
Signal invalid (may not be set in Peak Picker mode)
Invalid 1 Same as INV on rear panel
4
Case temperature too low
Same as Error Code X106
8
Case temperature too high
Same as Error Code X105
16
Detector block cold
Same as Error Code X104
32
Detector block hot
Same as Error Code X103
64
Current loop fault
Same as Error Code X108
128
Dirty window detection
Same as Error Code X101
256
Sensor failure
Same as Error Code X107 or FAIL on rear panel
display
512
DWD failure
Same as Error Code X109
1024
Signal invalid (Always set regardless of Peak
Picker mode). Measured Signal too low for ratio
measurement, Displayed Temperature is Peak
Picked or Hold mode Temperature
Invalid 2 Same as Error Code pINV on rear panel
display when Alarm button pushed
2048
RS485 Communications Lock
RS485 Communications Locked out, In read only
4096
Temperature reading under range
Same as LOW on rear panel TEMP display
8192
Temperature reading over range
Same as HIGH on rear panel TEMP display
16384
Laser pointer is on
Same as Laser LED on Sensor rear panel
32768
Unit under Cal test
Same as CAL on rear panel TEMP display