EasyManua.ls Logo

ircon Modline 5 - Determining Response Time to Use

ircon Modline 5
156 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Applications Guide
132
Rev. L6 Feb 2017 Modline 5
8.3.2 Determining Response Time to Use
The Response Time is factory set for the fastest response time allowed for its Model. You may wish to
select a slower response time that is more suitable for the process.
To follow temperature variations as they occur, select a fast response time. If you prefer to "filter out"
temperature variations, select a slower response time. In most applications, the practical approach is to
observe the temperature display or analog current output and adjust the response time for the most
meaningful temperature information without distracting variations. Observe the temperature
indications on a Temperature Display, or chart recorder. Figure 80 illustrates the general effects of
response times as if plotted by a chart recorder.
Figure 80: Effects of Response Time
Fast Response
Temperature
Time
Medium Response
Temperature
Time
Slow Response
Temperature
Time
Temperature display and output
signals follow temperature
variations as fast as minimum
response time of instrument.
Rapid temperature variations are
filtered out and system follows
general trends in temperature,
with small amounts of ripple.
Sensor tracks slow variations in
temperature ans effects of
temperature spikes are
eliminated.

Table of Contents

Other manuals for ircon Modline 5

Related product manuals