Technical Specification
Parameter IT6515C Ver:V1.6
Readback stability-8h
(%of Output +Offset)
Voltage
≤0.03%+100mV
Current ≤0.1%+30mA
Efficiency 90%(150V/20A)~93%(500V/6A)
Remote Sense
Compensation Voltage
5V
Command Response
Time
20mS
Power Factor 0.99
Ma Maximum input
current
8
10A
Maximum input
apparent power
2100VA
Storage temperature
-10℃~70℃
Protective function
Source OVP, OCP, OPP and Load OCP, OPP, OTP,Vsense
standard Interface USB/RS232/CAN/LAN
Isolation ( output to
ground)
500V
Parallel Number ≤8
Working temperature
0~40℃
Dimension ( mm)
483mmW×105.4mmH×640.8mmD
Weight( net) 17Kg
The above specifications may be subject to change without prior notice.
1. Internal absorption current and internal absorption power mean the
absorption current and power of the standard power dissipater inside the power
supply.
2. Resistance programming accuracy differs based on different output current.
For example, 10A Transient 1.0Ω, accuracy: (1.0Ω * 0.25%) +(375mΩ*A /10A)
= 40mΩ.1.0Ω, accuracy: (1.0Ω * 0.25%) +(375mΩ*A /10A) = 40mΩ.
3. Set value accuracy means the set accuracy realized through the panel button
or communication instructions. When external analog is used for programming,
the programming accuracy is 2%FS.
4. Read-back value accuracy means the read-back accuracy realized through
panel display or communication instructions; when external analog is used for
monitoring, the monitoring accuracy is 1%FS.
Copyright © Itech Electronic Co., Ltd.
119