Quick Start
 
  Copyright © ITECH Electronic Co., Ltd.  18 
Model: IT88XX 
Ver.: 1.XX-1.XX 
SN: XXXXXXXXXXXXXXXXXX 
Exception handling 
If the electronic load cannot start normally, please check and take measures by 
reference to steps below. 
1.  Check whether the power line is correctly connected and confirm whether 
the electronic load is powered. 
Correct wiring of power line => 2 
Incorrect wiring of power line => Re-connect the power line and check 
whether the exception is removed. 
2.  Check whether the power in On. Power key is under “ ” On status. 
Yes => 3 
No => Please check the Power key to start power and check whether the 
exception is removed. 
3.  Check whether set power voltage of electronic load is larger than the power 
supply voltage. If set power voltage is 220 V and the supply voltage is 110V, 
the electronic load cannot start. 
4.  Check whether the fuse of electronic fuse is burned out. 
If yes, change fuse. Detailed steps: 
1)  Pull out power line and take out the fuse box at power line jack with a 
small screw driver. As shown below. 
   
 
2)  If the fuse is fused, please change fuse of same specification based on 
machine model. See the table blow for matching information of fuse 
and machine model. 
Specification (110VAC)  Specification (220VAC) 
IT8813/13B/13C  T 3.15A       250VAC  T 1.6A   250VAC 
IT8814/14B/14C  T 3.15A  250VAC  T 1.6A   250VAC 
IT8816/16B/16C  T 3.15A  250VAC  T 1.6A   250VAC 
IT8817/17B/17C  T 5A     250VAC  T 2.5A   250VAC 
IT8818/18B  T 5A   250VAC  T 2.5A  250VAC 
IT8818C/18D  T 5A   250VAC  T 2.5A   250VAC 
IT8831/31B/31H  T 6.30A   250VAC  T 3.15A   250VAC 
IT8832/32B/32H  T 6.30A   250VAC  T 3.15A   250VAC 
IT8833/33B/33H  T 6.30A     250VAC  T 3.15A   250VAC 
IT8834B/34H  T 6.30A        250VAC  T 3.15A   250VAC