Function and Features
 
  Copyright © ITECH Electronic Co., Ltd.  24 
Constant  Resistance 
High=130.0V 
4.  Set minimum voltage value and press [Enter] key for confirmation. 
Constant  Resistance 
Low=0.000V 
5.  Complete parameter setting. 
10.0000V  0.0000A 
0.00W  CR=2.000Ω 
 
NOTE 
If the above method is for editing auto test step (as mentioned below), constant 
resistance range can also be set. 
3.2.4 Constant power operation mode (CW) 
Under CW mode, the electronic load will consume a constant power, as shown 
below. If input voltage rises, the input current decreases and power P (= V * I) 
will maintain at setting power. As shown in Fig. 3-4: 
 
 
Fig. 3-4 Voltage-Current Relation Schema under CW Mode 
 
Under CW mode, the electronic load provides three ways to modify constant 
power. 
  Under CW mode, rotate the pulse knob to change constant power value. 
  Under  CW  mode, directly input numeric key and press [Enter]  key for 
confirmation to change constant power value. 
  Under CW mode, move the cursor with Left/Right Key and press Left/Right 
Key to adjust value at corresponding position. 
Operation steps 
1.  Press [CW] key. Press  [Shift] + [CV] to enter parameter setting screen. 
Constant   Power 
Range=400.00W 
2.  Set maximum working power value and press [Enter] key. 
Constant   Power 
Input 
voltage 
Load current 
V
I
V3 
I3