Function and Features
 
  Copyright © ITECH Electronic Co., Ltd.  22 
Constant  Current 
Rise down=0.000A/uS 
8.  Complete parameter setting. 
10.0000V  0.0000A 
0.00W  CC=1.000A 
 
NOTE 
If the above method is for editing auto test step (as mentioned below), constant current 
range can also be set. 
3.2.2 Constant voltage operation mode (CV) 
Under CV mode, the electronic load will consume sufficient current to maintain 
the input voltage at setting voltage. As shown in Fig. 3-2. 
 
 
Fig. 3-2 Voltage-Current Relation Schema under CV Mode 
 
Under CV mode, the electronic load provides three ways to modify constant 
voltage. 
  Under CV mode, rotate the pulse knob to change constant voltage value. 
  Under CV mode, directly input numeric key and press [Enter]  key for 
confirmation to change constant voltage value. 
  Under CV mode, move the cursor with Left/Right Key and press Left/Right 
Key to adjust value at corresponding position.
 
Operation steps 
1.  Press [CV] key and [Shift] + [CV] to enter parameter setting screen. 
Constant  Voltage 
Range=120.00V 
2.  Set maximum working voltage value and press [Enter] key for 
confirmation. 
Constant  Voltage 
Range=2.33V 
3.  Set maximum current value and press [Enter] key for confirmation. 
Constant  Voltage 
High=66.000A 
Set current
Load 
  voltage
I 
V
Set voltage