Function and Features
 
  Copyright © ITECH Electronic Co., Ltd.  30 
screen status is on). 
TRANSITION 
Continuous  Pulse  Toggle 
3.  Operate Left/Right key and select high rate and low rate. Move to the 
High-Rate and press [Enter] key. 
TRANSITION 
High-Rate  Low-Rate 
4.  Set ascending slope and press [Enter] key. 
TRANSITION 
Rise up=2.000A/uS 
5.  Set descending slope and press [Enter] key for confirmation. 
TRANSITION 
Rise down=2.000A/uS 
6.  Set A value and press [Enter] key for confirmation. 
TRANSITION 
Level A=1A 
7.  Set B value and press [Enter] key for confirmation. 
TRANSITION 
Level B=2A 
8.  Set frequency value and press [Enter] key for confirmation. 
TRANSITION 
Frequency=50Hz (0.01-25000Hz) 
9.  Set duty ratio and press [Enter] key for confirmation. 
TRANSITION 
Duty=98% (0.1%-99.9%) 
10.  Start dynamic test and operate Left/Right key. Move to On and press 
[Enter] key for confirmation. 
TRANSITION 
On  Off 
11.  Enter the dynamic test mode. 
10.0000V  0.0000A 
0.00W   0 TRAN 
12.  Press [On/Off] key and press [Trig] key. 
The load will continuously switch between A and B value. Time of 
operations is shown at bottom right. 
13.  Press CC/CV/CR/CW key or any composite function key to exit dynamic 
test function. Repeat Steps 1-12 to continue parameter setting and 
operation of dynamic test. 
3.9.2 Pulse mode 
Under pulse mode, after enabling dynamic test operation, the load will switch to 
A value after receiving of a trigger signal. Then the load will switch back to B 
value and be constant at B value after maintaining A for pulse width time.