Function and Features
 
  Copyright © ITECH Electronic Co., Ltd.  40 
  RECALL 
1.  Press  [Recall]  key and press 9 (to select from which group the data is 
recalled). 
5.8949V   0.99994A 
5.89W    cc=1.000A 
Memory function 
When you want to recall the data saved in the memory, you should set memory 
group in the system menu first. 
Group 0 means you can recall data saved in 0-10 groups. Group 1 means you 
can recall data saved in 11-20 groups. Group2-Group 9 can be concluded in the 
same manner. 
3.16 VON function 
When testing some power products with slow voltage rise speed, if the 
electronic load input is opened before power, the power may latch protection. In 
this way, the user may set VON value. The electronic load only latches when 
power voltage is higher than this value. 
Press [Shift]+ 6 keys to enter configuration menu. Set voltage value in Voltage 
on under configuration menu to control on/off status of electronic load. When 
VON LATCH function is started, it indicates that Von Point latches load status. 
NOTE 
Please confirm whether it is necessary to set loading voltage, a step that provides 
convenience for limiting working voltage value. If not necessary, do not set the loading 
voltage without authorization to prevent unnecessary trouble from failure of loading. 
If the instrument cannot load, please firstly check whether the VON function is set. If yes, 
reset the Von value to minimum value (which may be directly set as 0. If minimum voltage 
value of instrument is not 0, press 0 for confirmation and the menu will automatically set 
the value as minimum value). 
This function is invalid under CV mode. 
When VON LATCH function is started, the load starts load test only when the 
power voltage rises and is higher than Von Point loading voltage. When the 
power voltage drops and is lower than Von Point unloading voltage, the load will 
not unload.