Function and Features 
  Copyright © ITECH Electronic Co., Ltd.  19 
Measure voltage rise/drop time 
Set stopping voltage value 
CR-LED 
CR LED MODE 
Analog LED lamp function (under CR 
mode) 
Remote-Sense 
REMOTE SENSE STATE     
Remote sense compensation 
Start remote sense compensation function 
Stop remote sense compensation function 
Ext-Program 
External analog quantity function 
Start external 0-10 V analog quantity control function 
Stop external 0-10 V analog quantity control function 
 
3.8 Triggering function 
Triggering function is necessary to operate dynamic pulse output and list output. 
There are five triggering methods to trigger the DUT. 
Optional triggering sources of triggering function of electronic load comprise: 
  Key ([Trigger] key) trigger: when key trigger is valid, press [Shift] + 
 
key and the load will trigger an operation. 
  External trigger signal (TTL): TRIG on the rear board is the triggering 
input terminal. When external trigger signal method is effective, after 
applying a low pulse (> 10 us), the load will trigger an operation. 
  Bus trigger: when the bus trigger is valid, after the load receives a trigger 
command (GET or * TRG) from GPIB port, the load will trigger an 
operation. 
  Timing trigger: when the timing trigger is valid, the load will automatically 
trigger an operation at intervals. 
  Trigger holding: when trigger holding is valid, only when the load receives 
trigger command (TRIG:IMM), the load will trigger an operation. 
  Select the triggering sources as follows: 
Operation steps 
1.  Press [Shift] + [8] (system) to enter system menu setting screen. 
Initialize Power-ON Buzzer 
2.  Press 
  to move to the Trigger and press [Enter] to enter 
triggering source selection screen. 
SOURCE MANUAL 
3.  Press 
  to select trigger method and press [Enter] to complete 
setting. 
Manual (Def): Manual trigger 
External: external signal trigger 
Hold: special command trigger 
Bus: BUS command trigger 
Timer: timer trigger 
4.  Press [Esc] to exit setting. The system displays original values under