Function and Features 
  Copyright © ITECH Electronic Co., Ltd.  33 
that indicates VFD screen status is on). Press [ESC] key to exit setting. 
LIST 
On  Recall  Edit 
4.  Press [On/Off] key to open input and press [Shift] + 
  key (Triggering 
key) 
List operation. 
3.19 Terminal function of rear board 
Terminals on IT8800 rear board comprise remote sensor terminal, external 
trigger terminal, external analog control terminal, voltage fault indication 
terminal, external On/Off control terminal and current monitoring terminal. 
Terminal schematic (as shown below): 
 
 
External analog control terminal 
Negative input terminal for TRIG, ON and VF   
Positive input terminal for trigger 
Positive input terminal for external On/Off control 
Positive input terminal for voltage fault indication 
 
3.19.1 Remote sense compensation functions 
Under CC, CV, CR or CP mode, if the load consumes large current, a large 
voltage drop will be detected in connection line between tested instrument and 
load terminal. To ensure measurement accuracy, a remote sense measurement 
terminal is provided at load rear board to compensate voltage drop lost in wire. 
Remote sense operation: SENSE (+) and SENSE (-) are remote input terminals. 
To avoid voltage drop caused by long input wire of load, the remote sense test 
allows direct measurement at input terminal source so as to improve 
measurement accuracy. 
Operation steps of remote sense measurement: 
1.  Press [Shift] + [9] keys to enter menu. Operate 
  key and select 
Remote-Sense. Press [Enter] key. 
2.  Select ON and start Sense function. Set load in remote sense 
measurement mode. 
3.  Remote sense measurement is connected. Refer to figure below for 
detailed wiring.