Function and Features 
  Copyright © ITECH Electronic Co., Ltd.  31 
When load temperature is decreased to protection point, press any key on load 
front board (or send command to PROTection:CLEar), the load front board 
(OTP) word will get cleared and the load will exit OTP protection status. 
3.18 List operation 
LIST mode provides an accurate, fast and low-cost way to complete any 
complicated current change mode, which enables synchronization of internal or 
external signals in multiple quasi-bit load precision tests. 
When different trigger sources are selected, the LIST function will form a variety 
of complex sequences by editing step value, pulse width and slope of each step 
to meet complicated test requirements. LIST parameters comprise designation 
of input list file, input step count (2-84 steps at maximum), step time (0.00002s 
– 3600s) as well as setting value and slope of each step. The list file can be 
stored in non-volatile RAM available for a quick output in case of usage. The 
user can edit 7 groups of list files at maximum. 
If the load operation mode is at List operation, the load will start List operation 
when it receives a trigger signal till completion or receipt of another trigger 
signal. 
Before List operation, be sure to edit List operation files and save them in load 
non-volatile RAM. Refer to examples below to know how to execute List 
operation through board. It is assumed that output voltage and current of the 
tested instrument are 10V and 3A respectively and the load is under CC mode. 
Trigger
10 2 3 4 5
List count=1 List count=2
List sequence
 
 
Edit the LIST file and trigger to operate this file. Operation steps: 
Operation steps 
1.  Press [Shift] + [3] key. Operate Left/Right key. Move to Edit and press 
[Enter] key. 
LIST 
On  Recall  Edit 
2.  Operate Left/Right key and move to the High-Rate and press [Enter] key. 
EDIT  LIST 
High-Rate  Low-Rate 
3.  Set CC range and press [Enter] key. 
EDIT  LIST 
Current Range=3A 
4.  Edit number of steps. Press [2] key to edit two steps. Press [Enter] key. 
EDIT  LIST 
File Step=2(2-84)