Function and Features 
  Copyright © ITECH Electronic Co., Ltd.  25 
If the set OCP voltage value is higher than the power voltage value, the OCP will fail 
to operate and the board will display as follows: 
 
9.9990V      0.0009A    
0.01W    0.100A    FAULT   STOP  
 
 
3.11 OPP test function 
The IT8800 series electronic load is provided with overpower protection test 
function (OPP). Under OPP test mode, when input voltage reached Von value, 
delay for a while for the electronic load to latch. Ascend value by step value at 
regular interval. At the same time, check the load input voltage and judge 
whether it is higher than standard voltage value. If higher, it indicates that OPP 
does not occur. Repeat power stepping operation till the load operates to the 
cutoff power; if lower, it indicates that OPP does occur. Check whether the 
existing power value is within target scope again. If yes, PASS the test. 
Press [Shift] + [CW] keys to enter OPP test function setting screen: 
OPP 
TEST 
Run 
Operate OPP test documents 
Recall 
Recall OPP test documents (1-5) 
Edit 
1:Voltage on level=0.000V 
Set Von voltage delay time 
9:Max Trip Power =0.000W  Set overpower range 
10:Min Trip Power =0.000W 
Set overpower range 
(minimum value) 
(
)
 
Operating steps: 
1.  Press [Shift] +
    key to start OPP test. If within range, PASS the test 
and the board will display as follows: 
 
9.996V         0.0007A 
0.01W    49.10W   PASS   STOP 
 
If not, there is FAULT and the board will display as follows: 
 
9.996V         0.0007A 
0.01W   48.6W   FAULT   STOP 
 
2.  End test. Press [ESC] to return to setting screen. Press [ESC] again to exit. 
NOTE