2. Select Maintain > Sensor Maintenance.
3. Clamp the Intelis Gas Meter to the prover and turn on appropriate nozzles in the center
column of the Sensor Maintenance window to exercise the meter.
Note: Nozzle 192 CFH is typically used to exercise the Intelis Gas Meter.
4. Exercise the meter for 10 to 60 minutes until the Out Temp value does not vary by more than
0.2° F within a minute.
5. Close the nozzle.
6. Unclamp the meter.
7. Exit the Sensor Maintenance window by clicking Done in the lower right corner of the win-
dow.
11 December 2019 TDC-1782-005 59
Intelis Gas Meter Installation Guide 5 Proving