The meter uses the entered values to calculate the required calibration parameters to a tar-
get of one of the following, depending on your selected prover station mode:
–
% Error. 0% open, 0% check
–
% Accuracy. 100% open, 100% check
–
% Proof. 100% open, 100% check
11 December 2019 TDC-1782-005 73
Intelis Gas Meter Installation Guide 6 Adjusting the Calibration