Cleaning and Maintenance Work for the User
130
9.3.4 Cleaning the milling chamber
CAUTION! Skin contact with the coolant fluid. Residues enter the water circuit.
Skin irritation.
Danger to the environment due to leaking coolant fluid.
When handling the coolant liquid and components which could contain the fluid (tank,
collection container, etc.), wear gloves.
Dispose of deposits of the milling chamber and collection container according to local
regulations.
Observe the Safety Data Sheet of the Coolant Concentrate PrograMill Fluid for One and
the documentation about the used blocks, “Supporting documents”, page 9.
NOTICE! Improper cleaning of the milling chamber with unsuitable equipment and
cleaning agents.
Damage to the membrane and other sensitive components in the inner chamber.
Do not clean with or enter the milling chamber using brushes, tools, sponges with metal
components, or any cleaning equipment other than that specified in this chapter.
Do not use sharp, chlorine-containing cleaners. Only use mild, pH neutral cleaning agents.
To avoid damaging the sensitive components in the milling chamber, avoid strong
pressure, scrubbing, rubbing and wiping.
In general, do not clean the processing and changer arm, the changer fork and nozzle
holes in the interior.
Do not move, relocate or manipulate the cover between the milling chamber and the
material and tool changer.
To avoid scratching and therefore clouding the window of the milling chamber, clean the
milling chamber only with the recommended cleaning accessories specified in this section
or in the app.
Do not wear jewellery such as rings, bracelets or watches when working on the device,
especially when cleaning the milling chamber.
NOTICE! Contact with the membrane when removing the collection container and cover.
Damage of the membrane.
Do not touch the membrane with the collection container or cover.