Programming Guide 64
Command
Parameters
Response
Unit
MIN
MAX
DEF
Command
Status
(table continued)
:OFFSet?
MAX | MIN | DEF
NRf
dB
0.00
99.99
0
SC
:WAVelength
Numeric Value
M
1200NM
1700NM
1310NM
N
:WAVelength?
MAX | MIN | DEF
NRf
M
1200NM
1700NM
1310NM
N
:OUTPut
[:STATe]
Boolean
N
[:STATe?]
NRf
N
:APOWeron
Boolean| DIS|
LAST
N
:APOWeron?
NRf
N
:APMode
Boolean
N
:APMode?
NRf
N
:DRIVer
Boolean
N
:DRIVer?
NRf
N
:POWer
Numeric Value
dBM
0.00
7
100.00
8
0.00
N
:POWer?
MAX | MIN | DEF
NRf
dBM
0.00
8
100.00
8
0.00
N
:STATus
:OPERation
[:EVENt]?
NRf
SC
:CONDition?
NRf
SC
:ENABle
NRf
0
32,76
?
SC
:ENABle?
NRf
SC
:NTRansition
NRf
0
32,76
SC
NTRansition?
NRf
SC
:PTRansition
NRf
0
32,76
SC
NRf
SC
7
:OUTP:POW sets the through power, and :OUTP:POW? returns the through power. The
maximum value for this command is the through power for which the actual attenuation is 100
or 60 dB, depending on the model. The minimum and default values are the through power,
for which the actual attenuation is 0 dB.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the JDS Uniphase HA9 Series and is the answer not in the manual?

JDS Uniphase HA9 Series Specifications

General IconGeneral
BrandJDS Uniphase
ModelHA9 Series
CategoryControl Unit
LanguageEnglish

Related product manuals