Programming Guide – 36
HA9/HPM Query Commands
D?
Beam Block Status
Returns the on/off status of the beam block:
• 0 = beam block off
• 1 = beam block on
WVL?
Calibration Wavelength
Returns the calibration wavelength:
• WVL? returns the current calibration wavelength
• WVL? MIN returns the minimum calibration wavelength
• WVL? MAX returns the maximum calibration wavelength
SLP?
Slope User Value
Returns the user slope:
• SLP? returns the current user slope
• SLP? MIN returns the minimum user slope
• SLP? MAX return the maximum user slope
USER?
Slope Control
Returns the method by which the attenuator determines the slope of the attenuation:
• 0 = calibration wavelength
• 1 = user slope
DISP?
Display Mode
Returns the display mode of the attenuator:
• 0 = ATT mode
• 1 = PWR mode