Programming Guide – 63
-430
Query deadlocked
The device was deadlocked. Both the input buffer and output queue are full, and the
attenuator is unable to continue.
??
Undefined error number
An undefined error occurred.
SCPI Command Tree
All commands other than the IEEE 488.2 common commands are listed in Table 9. The
following abbreviations are used in the Command Status column:
• SC (SCPI-confirmed command)
• SA (SCPI-approved command)
• N (command is neither SCPI confirmed nor approved)
Table 9: SCPI Commands
Command
Parameters
Response
Unit
MIN
MAX
DEF
Command
Status
:DISPlay
:BRIGhtness
Numeric Value
0
1
1
SC
:BRIGhtness?
MAX | MIN | DEF
1
0
1
1
SC
:ENABle
Boolean
SC
:ENABle?
NRf
SC
:INPut
:ATTenuation
Numeric Value
dB
0.00
6
100.00
7
0
SC
:ATTenuation?
MAX | MIN | DEF
NRf
dB
0.00
7
100.00
7
0
SC
:LCMode
Boolean
N
:LCMode?
NRf
N
:MINLoss
N
:OFFSet
Numeric Value
dB
0.00
99.99
0
SC
:DISPlay
N
6
:INP:ATT sets the total attenuation, and :INP:ATT? returns the total attenuation. Therefore,
the maximum value for this command and query is the current offset (100 or 60 dB depending
on the model). The minimum and default values are the current offset 0.dB.