Programming Guide 80
System Commands
:SYSTem:ERRor?
Syntax
:SYSTem:ERRor?
Function
Returns the next error message in the error queue. Because the error queue is
an FIFO queue, the error returned is the oldest unread error. The error message
consists of an error number followed by an error message, for example,
0, No Error
See the Error Queue section for a list of error numbers and their associated
messages.
Example
:SYST:ERR?
:SYSTem:VERSion?
Syntax
:SYSTem:VERSion?
Function
Returns the formatted numeric value corresponding to the SCPI version number
to which the attenuator complies.
Example
:SYST:VERS? returns 1995.0
User Commands
:UCALibration:USRMode
Syntax
:UCALibration:USRMode <space> <boolean>
Function
Sets user mode on or off. A boolean value of 1 or ON turns user mode on and
the HA9 uses the current user slope instead of the factory-set slope.
A boolean value of 0 or OFF turns user mode off and the HA9 uses the factory-
set slope.
Example
:UCAL:USRM ON sets user mode to ON.
:UCALibration:USRMode?
Syntax
:UCALibration:USRMode?
Function
Returns the current setting of user mode, that is, returns 1 if user mode is on and
0 if user mode is off.
Example
:UCAL:USRM OFF;USRM? returns 0.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the JDS Uniphase HA9 Series and is the answer not in the manual?

JDS Uniphase HA9 Series Specifications

General IconGeneral
BrandJDS Uniphase
ModelHA9 Series
CategoryControl Unit
LanguageEnglish

Related product manuals