Parameter setup and measuring operation
LDM52 OEM, Release 003, Revision 000
6.4 Status
6.4.1 AS – Autostart
The autostart function defines the behavior of the LDM52 OEM after a cold boot.
After the connection to the supply voltage and the internal start-up routine the LDM52 OEM will
automatically execute the command and send the data to the available outputs.
A figure from the table below must be entered.
The output shows the command.
1 … 6, 13 … 18 (see table below)
Output: Autostart commands [AS]: DT
Depending on the measurement mode used, it takes max. 6 sec from applying the supply
voltage to the point where the first measured value is put out.
Parameter x see table below
Output of device identification
Output of internal device temperature
Start of individual measurement
Start of continuous measurement
Start of quick continuous measurement
Following parameters are usable if pin 16/ HEIZ_EN (controller output of external heater) is connected
Heating is deactivated + output of device identification
Heating is deactivated + output of internal device temperature
Heating is deactivated + start of individual measurement
Heating is deactivated + start of continuous measurement
Heating is deactivated + start of uninterrupted continuous measurement